Controller
|
Description |
Control
Functions: |
Dual
PID+: two independent, event-selectable sets of control functions for each channel consisting of proportional, integral and derivative
control plus reset windup inhibit and reset clipping. "Set
2" tuning is in effect when the enabling event is ON. Cycle
time ranges from 2 to 15 seconds. |
Time
Proportioning: |
Two
per channel (increase & decrease); clamp logic outputs for driving
photo-isolated solid state relay. |
Linear
Outputs: |
Two
outputs, each assignable to either channel as process variable or
setpoint retransmit. 4-20mA or 0-16mA; maximum loop resistance,
600 Ohms. |
Cold
Junction Comp: |
±0.01
degree/ degree over 15°C to 35°C ambient operating range.
|
Physical
Characteristics |
Description |
Overall
Dimensions: |
6"
x 6" x 12 7/8" (w/o accessories); 10 lbs. |
Mounting: |
5
3/8" square cutout. |
Power
Requirements: |
117VAC
±10%; 25W. |
Operating Temperature: |
0
to 40C; 85% max. RH, non-condensing. |
Alphanumeric Displays: |
two
2-line, 20 character per line, Vacuum Fluorescent Display. |
Setpoint
Programmer |
Description |
Setpoint
Range: |
Full
range of the controller. |
Program
Steps: |
200
total steps; 99 programs, max. |
Step Types: |
RAMP/SOAK
(Standard), EOP (End of Program); LOOP; GOTO; GOTO_IF and PAUSE. |
Step Time Duration: |
0
to 99:59:59. (HH:MM:SS) |
Delay
Start: |
0 to 99:59:59. (HH:MM:SS) |
Guaranteed
Soak: |
Program
waits at the end of a ramp step until the process is within specified
tolerance; programmable, per step, in 1 unit increments, per channel. |
External
Events: |
16
programmable clamp logic outputs suitable for driving photo-isolated
solid state relays. |
Process Deviation Alarms: |
±1
to 55 units, programmable per channel, per step. |
Display
Deviation Alarm Action: |
Flashing indication on front panel display plus clamp logic output;
programmable to clear automatically or to require manual reset. |
Deviation
Alarm Action: |
Programmable
to continue running; or pause until condition clears or until alarm
is manually reset. |
HI/LO
Process Limits: |
Programmable
per channel in 1 unit increments over full input range. |
Limit
Outputs: |
One
each per channel, logic output suitable for driving photo-isolated
solid state relay. |
Control
Inputs: |
External
photo-isolated inputs provided for run, stop, reset and front panel
lockout. |
Recovery
After Power Failure: |
Three
power fail recovery modes: Normal (auto resumption of program);
Hold (program and setpoint values at time of power fail); Run (selected
program and step #). |
Accuracies are
given for an ambient temperature of 23°C ±2°C. Thermocouple
tables 1968.
1.
Resolution for all inputs from -99.0 to 999.9 = 0.1 and for below
-99.9 and above 999.9 = 1.0.
2.
100 Ohm platinum, a=0.00385.
3.
Common positive current ranges (e.g. 4-20 ma, 0-16 ma)
4.
Common positive voltage ranges (e.g. 0-1 V, 0-5 V, 0-10 V)
5.
Please see manufacturer's specification tables.
6.
The display will show 0.0 to 100.0 for 0.0 to 100,000 ft.
The JC Systems family of Precision Environmental
Chamber Programmer/Controllers puts it all together for you: upward
compatibility, configuration flexibility, ease of operation, convenient
packaging plus the dedication to performance, reliability and service
you've come to expect from JC Systems - the leader in chamber control
innovation.